Last updated on Oct 28, 2020
-
Reanalyzing an analysis while the Assay Plugin has been inactivated, results in a Failed analysis. Reactivate the relevant Assay plugin prior to reanalyzing or upload the file again so that the new Assay plugin will be chosen.
-
Usability issues:
-
Resetting the scale of amplification graph settings does not take the x-axis into account.
-
When changing a result, the audit trail mentions Positive and Negative instead of Detected and Not detected
-
Unsupported file formats (e.g. .pdf) can be selected during the first analysis
-
Difficulties when selecting a single curve in an amplification graph
-
Collapse/Expand all rows in the Summary tab may not always work
-
Selection is not remembered in the plate and graph view after editing a result
-
Deselecting wells in plate view does not deselect the corresponding curves
-
Resetting graph zoom when using logarithmic scale fails
-
Selecting wells in the results plate view via dragging does not work due to a scaled plate view
-
Opening an analysis may return an error ("Cannot read property 'x' of undefined") due to failed loading of fluorescence values. Refreshing the page solves the issue.
-
Comments
0 comments